Product Model |
TRIX-16. |
Sensor Measurement Range |
-40°C to +85°C (-40°F to +185°F) |
Operating Temperature Range |
-40°C to +85°C (-40°F to +185°F). |
Storage Temperature Range |
-30°C to +55°C (-22°F to +131°F). |
Rated Temperature Reading Accuracy |
Better than ±0.5°C for -20°C to +40°C.
Better than ±0.7°C for -30°C to -20°C and +40°C to +60°C.
Better than ±0.8°C for +60°C to +85°C.
Better than ±0.9°C for -40°C to -30°C.
Actual performance is typically much better than the rated values. Please see the Rated Absolute Accuracy chart below.
Accuracy figures can be improved by recalibration. |
Rated Temperature Reading Resolution |
Less than 0.1°C for -40°C to +40°C.
Less than 0.2°C for +40°C to +80°C.
Less than 0.4°C for +80°C to +85°C.
Please see the Rated Native Resolution chart below.
LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®. |
Sensor Reaction Time |
Typically less than 5 minutes (T90) in moving air (1m/s). |
Recording Capacity |
16,159 temperature readings. 112 days @ 10min logging, 168 days @ 15min logging. |
Sampling Interval |
Configurable from 30 seconds to several hours. |
Logging Start Options |
Push button start or specific date & time. Optional start delay of up to 18hours. |
Recording Indication |
Flashing ‘OK’ indicator / flashing ‘ALERT’ indicator. |
Download Time |
Typically less than 10 seconds for full memory (16,159 readings), depending on computer or readout device used. |
Environmental |
IP65 (roughly equivalent to NEMA 4). |
Power Source |
3V Li-Mg Battery. |
Battery Life |
2 – 3 years of normal use (based on 15 minute logging, download data monthly). |
Real Time Clock |
Built-in real time clock.
Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day).
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C). |
Size |
86mm(H) x 54.5mm(W) x 8.6mm(T). |
Weight |
33g. |
Case Material |
Polycarbonate. |
Compliance and Certifications |
|